The IMMULITE® 2000 XPi System is an easy to use, continuous random-access immunoassay analyser with one of the largest automated immunoassay menus available.
With a maximum throughput of 200 tests per hour, medium to high-volume laboratories can incorporate speciality and allergy testing into routine workloads to improve productivity and efficiency.
Features
- Customised automated startup for daily routine maintenance and quality control scheduling
- Reagent onboard stability of 90 days, minimises waste
- Maximum throughput of 200 tests per hour
- No pause sample loading to enhance turnaround time
- Tube-top sample cups for micro-sample handling
- Home screens with at-a-glance status on a 19" touch screen monitor
- Options available for direct water feed and bulk solid waste chute
- Siemens Automation solution available through VersaCell® Systems
Specifications
Overview |
System Description
| Random access immunoassay system
|
Throughput
| Up to 200 tests per hour in batch or random access mode
|
Time to First Result
| 35 minutes
|
Assays Onboard
| 24
|
Assays |
Assay Onboard Stability
| 90 days
|
Assay Technology
| Enzyme amplified Chemiluminescence
|
Disease-state Assay Groups
| Allergy, Anaemia, Autoimmune, Bone, Cardiac, Diabetes, Growth, Hepatitis, Inflammation, Metabolic, Oncology, Reproductive, Special ID, Therapeutic Drug Monitoring, Thyroid, ToRCH, Veterinary
|
Assay Calibration
| System indicates when calibration is required Specific sample type available for calibration verification
|
Samples and Sample Handling |
Sample Tubes
| 12-16mm diameter, 75-100mm height, 10 x 50 microtubes with tube top racks
|
Sample Carriers
| 90 samples, continuous loading in 15 position sample racks
|
Walk-away Time
| Up to 5 hours
|
Sample Types
| Serum, plasma, urine - assay dependent
|
Sample Integrity Control
| Pressure-transducer technology with clot detection, clot flagging and short sample detection
|
Automatic Processing
| Reflex and re-run testing and dilutions
|
Barcodes |
Formats Supported
| Code 39, Codabar, UPC, Code 128, 1-2 of 5 (interleaved)
|
Reagents and Reagent Handling |
Reagent Cooling
| 2-8°C
|
Reagent Preparation
| None required
|
Reagent Onboard Stability
| 90 days
|
Reagent Integrity Control
| Barcodes allow system to identify lot number, determine when calibrator adjustment needed and the remaining number of tests in the reagent wedge
|
Cuvettes |
Type
| Individual reaction tubes
|
Onboard Capacity
| 1300
|
Maintenance |
Daily
| Automated daily maintenance cycle completed according to time programmed by the laboratory
|
Weekly
| 20 minutes
|
Monthly
| 45 minutes
|
User Interface |
Operating System
| Microsoft® Windows® 7
|
Monitor
| 19" touch sensitive screen
|
Quality Control |
Package
| Includes graphs and QC rules options for controls
|
Environmental |
Operating Temperature
| 18-32°C and 40-80% humidity
|
Electrical
| 200-240 V, 6 AMP, 50/60Hz, 6 Amp
|
Weight & Dimensions |
Weight
| 363kg
|
Height
| 165cm
|
Width
| 236cm
|
Depth
| 114cm
|