InSight® V-EL is a whole blood Veterinary Electrolyte Analyser that requires zero maintenance and can test 5 electrolytes on 1 single-use cartridge.
Features
- Tests electrolytes K+, Na+, Cl-, iCa2+ and iMg2+ in whole blood on 1 single-use cartridge
- Accurate results in 5 minutes
- Auto-calibration before each test
- 110µl whole blood sample volume
- Dry chemistry method, single-use cartridge eliminates the risk of contamination
- Maintenance free - no fluid pipe inside the device, no reagent pack required
- Integrated barcode reader and printer
- Portable
Performance
Electrolyte
| Accuracy
| Precision
| Stability
| Test Range
|
K+
| ±3.0%
| ≤1.5%
| ≤2.0%
| 1.0-15.0 mmol/L
|
Na+
| ±3.0%
| ≤1.5%
| ≤2.0%
| 100-200 mmol/L
|
Cl-
| ±3.0%
| ≤1.5%
| ≤2.0%
| 65-160 mmol/L
|
iCa2+
| ±5.0% or ±0.05 mmol/L
| ≤1.5%
| ≤3.0%
| 0.25-4.00 mmol/L
|
iMg2+
| ±5.0% or ±0.05 mmol/L
| ≤3.0%
| ≤3.0%
| 0.2-1.5 mmol/L
|
How to Use
|
|
|
1. Scan the barcode on the cartridge pouch and take the cartridge out of the pouch.
| 2. Fill the inlet with the sample up to the fill mark and slide the cap to seal the inlet.
| 3. Insert the cartridge into the analyser until it clicks. Wait for the results.
|
Specifications
Sample Type
| Venous whole blood
|
Testing Principle
| Dry Electrochemistry/Microfluidic Method
|
Sample Size
| 110µl
|
Testing Time
| 5±0.5 min (from test cartridge insertion to result output)
|
Data Capacity
| 99,999 pieces of data (including test and QC results)
|
Usage Temperature
| 10°C~30°C
|
Display Screen
| 7" colour touch screen
|
Printer
| Built-in
|
Code Scanner
| Built-in
|
Dimensions
| 240 x 120 x 110 mm (± 10 mm)
|
Weight
| 1.5±0.2kg (with battery)
|
Ordering Information
Product Code | Product Description |
WD2300 | InSight V-EL Electrolyte Analyser |
WD2301 | InSight V-EL Test Cartridges (25) |
WD2302 | InSight V-EL Control Solution (6) (x2 High, x2 Normal, x2 Low) |
If you would like more information please contact Woodley Equipment on +44 (0) 1204 669033 or email sales@woodleyequipment.com